Symptom

The ranging result shows the radar reporting “within blind zone”.

Root cause

The working condition produced a weak reflection signal, so the returned data fell into the antenna blind zone.

Fix

Change the installation to suspended mounting (keep the antenna away from the surface and obstructions) and set the blind zone to 0.3 m. The issue was resolved.

Suspended installation
Suspended installation
Blind zone setting

At a glance: when the display shows “within blind zone” the reflection is too close to the antenna to measure. A weak reflection placed the data inside the blind zone; suspended mounting and a 0.3 m blind-zone setting resolve it.

Step-by-step diagnosis

  1. Read the displayed distance and compare it to the blind-zone value.
  2. Confirm the target is closer than the current blind-zone setting (default 0.3 m).
  3. Check whether the reflection is weak (low dielectric or oblique surface).
  4. Decide between re-mounting and adjusting the blind-zone parameter.

How to fix it

Switch to suspended mounting

Hanging the sensor below the bracket moves the surface reflection out of the near-field blind zone.

Set the blind zone to 0.3 m

Match the blind-zone parameter to the real minimum measuring distance so valid echoes are accepted.

Improve the reflection

For weak targets, increase RF power or reposition to a flatter surface.