Information Extractable From Echo Spectrum And Correspond...

Information extractable from echo spectrum and corresponding fixes

Spurious echoes detected → Enable spurious echo learning. Secondary echoes detected → Enable nearest peak detection and inspect RF settings. Weak/unstable true target echo energy → Optimize RF power and filter para

2026-08-07 
Empty Distance Data Fluctuation Analysis

Empty distance data fluctuation analysis

Scenario 1: Decreasing empty distance Fluctuation ≤ 0.4 m: Likely antenna echo interference. Fluctuation > 0.4 m: Obstacles exist between radar and target. Scenario 2: Increasing empty distance Readings are appr

2026-08-07 
Why Is There No Blind Zone Alert When The Target Falls Wi...

Why is there no blind zone alert when the target falls within the blind zone?

Answer: Suspended installation mode is enabled. Switch to standard installation mode to activate blind zone alerts (default blind zone 0.3 m, adjustable downwards). Suspended installation function: In complex echo scenar

2026-08-07 
Image7

How do I adjust the transmitting power of the device?

Answer: RF transmit power is normally left at the factory default — raising it is a fix for a specific symptom, not a general performance boost. When to raise it The true target echo is weak or unstable, or the de

2026-09-26 
Faq 3

Why does the 1.5-inch 80G radar show zeros or spike values in historical data?

Model: JJY-YLD128HS-1.5 — 1.5-inch 80 GHz radar level transmitter Symptom During commissioning the live readings were normal, but the historical data contained zeros. Also, when the actual level was 0.5 m, occ

2026-09-30 
Faq 1

Why does the short-range 80G radar read the wrong distance and throw error codes?

Model: JJY-YGD38M-18L — 80 GHz short-range radar level transmitter Symptom The customer needed the radar to measure correctly at an actual distance of about 0.12 m. With the radar lifted away from the target t

2026-09-30 
Faq 5

Why does the 80G radar with display report “within blind zone”?

Model: JJY-YLD128SP — 80 GHz radar level transmitter with display Symptom The ranging result shows the radar reporting “within blind zone”. Root cause The working condition produced a weak reflection si

2026-09-30 
Image3

Why does the M38 threaded radar read 1105 mm when the actual level is 510 mm inside a narrow pipe?

Answer: The radar was mounted off-center (or tilted) in the 108 mm pipe, so part of the beam hit the pipe wall and the sensor locked onto the wall echo — about 1105–1106 mm — instead of the tr

2026-09-23 
Why Does The Liquid Level Display Insufficient Echo Energy

Why does the liquid level display “Insufficient Echo Energy”?

Answer: Possible causes: Obstacles between the sensor and the target weaken reflected echo energy. The sensor is not parallel to the target surface, leading to weak reflected echoes. The measured target is not a flat sol

2026-08-07 
Image6

Why is the 1.5-inch 80G radar level reading inaccurate for grain, and which parameters fix it?

Answer: Bulk solids such as soybeans scatter the radar signal, so with default parameters the surface echo is misidentified and the height reads inaccurate. Field-proven tuning for the JJY-YLD128HS-1.5 on a grain silo: r

2026-09-23 
Measurement Shows Within Blind Zone

Measurement shows within blind zone

Answer: The measured distance is shorter than the blind zone distance; the factory default blind zone is 0.3 m. Solutions: Adjust the sensor-to-target distance to ensure it exceeds the blind zone. Reduce the blind zo

2026-08-07 
Can The Radar Be Used Directly After Installation

Can the radar be used directly after installation?

Answer: Additional configuration is required: set the installation height; calibrate current for 4‑20 mA output models. Installation height: Vertical distance from the water bottom to the radar hemisphere. After inputtin