Seedling Tray Level Reads 65535: A Target Inside the Blind Zone

You are here : Home > Radar & Ultrasonic Level Meter FAQ Hub > Radar Level Sensor Field Cases: Real Sites, Real Fixes > Seedling Tray Level Reads 65535: A Target Inside the Blind Zone
Industry: seedling propagation
Application: water level in a seedling tray
Device: JJY-YGD38M-18L radar level sensor
Symptom: level output jumps to 65535, then reads 0.14 m with 1 cm of empty distance left
Recorded: 5 November 2026

The problem

Site & symptom

The sensor sat 89 cm above a hydroponic seedling tray. The customer needed the whole range covered, including the last centimetre – the condition where the water surface comes almost up to the antenna.

Two failures appeared at that end of the range. With the water level close to the antenna, the reported level jumped to 65535. Moved closer still, with the water surface inside the blind zone, the unit returned an empty distance of 0.14 m instead of a level.

Root cause

The site photos showed the sensor mounted vertical, so the beam was not tilted – that was ruled out first.

The echo spectrum showed the target already inside the blind zone, while the unit was still outputting an empty-distance value. That is where the 65535 came from: with nothing valid to convert, the level output ran to its maximum instead of reporting the condition.

The device blind zone is 0.04 m. The lowest water position left only 1 cm of empty distance – a quarter of the blind zone. At that mounting height the last centimetre was never physically measurable, whatever the firmware did with it.

The installation

Radar level sensor mounted through the lid of a hydroponic seedling tray, surrounded by seedling plugs
Field case – the sensor mounted through the tray lid. The tray is only 89 cm deep, so the last few centimetres of water sit close to the antenna.
Close-up of the threaded mounting boss and cable gland where the sensor passes through the tray lid
The threaded mounting boss through the lid. The antenna face sits almost flush with the inside surface – nothing stands between the water and the sensor, and nothing protects it from being too close.

How we fixed it

Solution

Two changes, in this order:

(1) Firmware 20260424 → 20260901. On the older build the unit still produced an empty-distance value while the target sat inside the blind zone. After the upgrade it reports the blind zone as a blind zone instead of returning a false distance.

(2) Raised the sensor by 0.05 m. The lowest water position moved from 1 cm to 6 cm of empty distance – clear of the 4 cm blind zone, so the full range can be measured. The mounting height stored in the configuration was updated to the new value, not left at the old one.

Result

Every water level in the tray is now readable, and the reported level matches the actual level.

What the unit showed at the time

FFT echo spectrum with the target inside the blind zone, annotated to note that an empty distance value is still being output
Field capture – the target is inside the blind zone, and an empty-distance value is still coming out of the unit.
Configuration software showing a 1.00 m span, a 0.04 m blind zone, a live empty distance of 0.055 m and a level of 0.075 m
The same device in the configuration tool: 1.00 m span, 0.04 m blind zone. Live empty distance 0.055 m, level 0.075 m – both inside the last few centimetres of the range.

What this case teaches

  • Check the blind zone against the closest target, not the farthest. A 4 cm blind zone looks harmless on a 1 m span until you notice the water comes within 1 cm of the antenna.
  • 65535 is not noise. It means the unit has no valid echo to convert. Turning up the gain or re-learning the target will not clear it.
  • Firmware can stop the unit lying; it cannot make the last centimetre measurable. Only raising the antenna moves the closest target out of the blind zone.
  • Update the mounting height after you raise the sensor. If you add 5 cm to the antenna and leave the old value in the configuration, every reading is offset by 5 cm.

Measuring water level in a tray, sump or small tank where the surface comes close to the antenna? Send us the mounting height and the lowest level you need to see – we will tell you whether it falls inside the blind zone.