Why does the radar water level measurement lag and respond slowly?

Answer: Our 80G radar operates at 76‑81 GHz and outputs valid data 1 second after power-on. Amplitude limiting filtering is enabled by factory default to smooth readings. Disable filtering for faster response — measureme

2026-08-07 

Can the radar be used directly after installation?

Answer: Additional configuration is required: set the installation height; calibrate current for 4‑20 mA output models. Installation height: Vertical distance from the water bottom to the radar hemisphere. After inputtin

2026-08-07 

Why does the liquid level display “Insufficient Echo Energy”?

Answer: Possible causes: Obstacles between the sensor and the target weaken reflected echo energy. The sensor is not parallel to the target surface, leading to weak reflected echoes. The measured target is not a flat sol

2026-08-07 

Measurement shows within blind zone

Answer: The measured distance is shorter than the blind zone distance; the factory default blind zone is 0.3 m. Solutions: Adjust the sensor-to-target distance to ensure it exceeds the blind zone. Reduce the blind zo

2026-08-07 
  