No response when reading radar data via USB-to-serial adapter and upper computer software

supply voltage matches the device specification; abnormal voltage prevents normal operation. Check wiring: TTL communication: Cross wiring rule – RX ↔ TX. RS485 communication: Straight wiring rule – A to A, B to

2026-08-07 

Radar measurement on plastic disc targets

Operation guide: Cover the full disc surface with intact aluminum foil, no gaps or wrinkles. Add a waterproof film for long-term outdoor use to prevent aluminum foil oxidation and signal attenuation. For long-term

2026-08-07 

Measurement performance with heavy foam and floating debris

Answer: Foam thicker than 3‑4 cm triggers an “insufficient echo energy” alert. As foam thickness increases, radar signal penetration degrades, the sensor fails to capture valid target echoes, and outputs weak

2026-08-07 

How to output raw ADC data on the upper computer

Send AT command AT+OUTRAWDATA=2 via upper computer; confirm reply “OK”, then close the serial port. Open the Spectrum Viewer software, select the correct serial port, open the port, and click FFT Spectrum

2026-08-07 

Can multiple RS485 radars be wired in parallel? Can the upper computer read all units simultaneously?

Answer: Set a unique communication address for each radar (factory default address: 127 / Modbus hex address 0x7F). Radars with unique addresses can be connected on the same RS485 bus, but the upper computer software can

2026-08-07 

Can the ±2.75° beam angle be reduced? Minimum distance from silo wall for installation?

Answer: The beam angle cannot be reduced beyond its current specification — narrower angles increase interference from ground reflections and wall obstructions and require high-precision customized sensors with special c

2026-08-07 

Extending power supply cables

Q: Can I extend the power cable myself? Can the cable be extended to 50 m or more? A: Power cables for RS485 models may be extended on-site. Use thicker-gauge wiring to maintain operating voltage between 9‑12 V and avoid

2026-08-07 

What is the millimeter-wave operating frequency, modulation scheme, and supported communication protocols?

Answer: Operating frequency: 80 GHz; Modulation: FMCW (Frequency Modulated Continuous Wave); Supported protocols: Modbus, Text protocol.

2026-08-07 

Environment-specific tuning

1. Open environments (rivers, open water tanks) Typical issues: Weak true target echoes, frequent antenna echo interference, unstable distance readings. Increase RF transmit power (secondary echoes rarely occur in open

2026-08-07 

Information extractable from echo spectrum and corresponding fixes

Spurious echoes detected → Enable spurious echo learning. Secondary echoes detected → Enable nearest peak detection and inspect RF settings. Weak/unstable true target echo energy → Optimize RF power and filter para